The diagnostic accuracy of different fetal blood sample lactate cutoffs in labor, utilizing the StatStrip Xpress® lactate‐meter: A population‐based observational study

Abstract Introduction A recent study recommended 5.2 mmol/L as a cutoff for fetal blood sample (FBS) lactate in labor for the StatStrip Lactate®/Lactate Xpress® lactate meter. In the present study, we validated FBS lactate cutoffs in a larger population‐based setting, with different CTG guidelines,...