Evaluating small coronary stents with dual-source photon-counting computed tomography: effect of different scan modes on image quality and performance in a phantom

PURPOSE: The study aimed to assess the feasibility and image quality of dual-source photon-counting detector computed tomography (PCD-CT) in evaluating small-sized coronary artery stents with respect to different acquisition modes in a phantom model. METHODS: Utilizing a phantom setup mimicking the...