Amperometric acetylcholinesterase biosensor for physostigmine determination

A biosensor based on acetylcholinesterase immobilized by carbodiimide binding on glassy carbon electrode modified with carbon black and Co(II) phthalocyanine has been developed. The working conditions for assembling of the surface layer and measurement protocol have been specified for mediated oxida...

Full description

Saved in:
Bibliographic Details
Main Authors: R.R. Davletshina, A.N. Ivanov, G.A. Evtugyn
Format: Article
Language:English
Published: Kazan Federal University 2018-03-01
Series:Учёные записки Казанского университета: Серия Естественные науки
Subjects:
Online Access:https://kpfu.ru/amperometric-acetylcholinesterase-biosensor-for_340048.html
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1841564201793880064
author R.R. Davletshina
A.N. Ivanov
G.A. Evtugyn
author_facet R.R. Davletshina
A.N. Ivanov
G.A. Evtugyn
author_sort R.R. Davletshina
collection DOAJ
description A biosensor based on acetylcholinesterase immobilized by carbodiimide binding on glassy carbon electrode modified with carbon black and Co(II) phthalocyanine has been developed. The working conditions for assembling of the surface layer and measurement protocol have been specified for mediated oxidation of the product formed in the enzymatic hydrolysis of acetylthiocholine. It has been demonstrated that immobilization of the enzyme does not decrease the affinity of the active site to the substrate. The possibility for determination of physostigmine as a model reversible inhibitor in the concentration range from 0.3 nM to 10 μM and the detection limit of 1.8?10–10 M has been established. The kinetic parameters of enzymatic hydrolysis of the substrate and those of competitive inhibition of the enzyme have been calculated. The developed biosensor can find application for the monitoring of anticholinesterase drugs.
format Article
id doaj-art-d02b4a6fce9a46518d008203aa214462
institution Kabale University
issn 2542-064X
2500-218X
language English
publishDate 2018-03-01
publisher Kazan Federal University
record_format Article
series Учёные записки Казанского университета: Серия Естественные науки
spelling doaj-art-d02b4a6fce9a46518d008203aa2144622025-01-02T23:02:10ZengKazan Federal UniversityУчёные записки Казанского университета: Серия Естественные науки2542-064X2500-218X2018-03-011601516Amperometric acetylcholinesterase biosensor for physostigmine determinationR.R. Davletshina0A.N. Ivanov1G.A. Evtugyn2Kazan Federal University, Kazan, 420008 RussiaKazan Federal University, Kazan, 420008 RussiaKazan Federal University, Kazan, 420008 RussiaA biosensor based on acetylcholinesterase immobilized by carbodiimide binding on glassy carbon electrode modified with carbon black and Co(II) phthalocyanine has been developed. The working conditions for assembling of the surface layer and measurement protocol have been specified for mediated oxidation of the product formed in the enzymatic hydrolysis of acetylthiocholine. It has been demonstrated that immobilization of the enzyme does not decrease the affinity of the active site to the substrate. The possibility for determination of physostigmine as a model reversible inhibitor in the concentration range from 0.3 nM to 10 μM and the detection limit of 1.8?10–10 M has been established. The kinetic parameters of enzymatic hydrolysis of the substrate and those of competitive inhibition of the enzyme have been calculated. The developed biosensor can find application for the monitoring of anticholinesterase drugs.https://kpfu.ru/amperometric-acetylcholinesterase-biosensor-for_340048.htmlacetylcholinesteraseenzyme inhibitorbiosensordetermination of reversible inhibitioncobalt(ii) phthalocyaninechemically modified electrode
spellingShingle R.R. Davletshina
A.N. Ivanov
G.A. Evtugyn
Amperometric acetylcholinesterase biosensor for physostigmine determination
Учёные записки Казанского университета: Серия Естественные науки
acetylcholinesterase
enzyme inhibitor
biosensor
determination of reversible inhibition
cobalt(ii) phthalocyanine
chemically modified electrode
title Amperometric acetylcholinesterase biosensor for physostigmine determination
title_full Amperometric acetylcholinesterase biosensor for physostigmine determination
title_fullStr Amperometric acetylcholinesterase biosensor for physostigmine determination
title_full_unstemmed Amperometric acetylcholinesterase biosensor for physostigmine determination
title_short Amperometric acetylcholinesterase biosensor for physostigmine determination
title_sort amperometric acetylcholinesterase biosensor for physostigmine determination
topic acetylcholinesterase
enzyme inhibitor
biosensor
determination of reversible inhibition
cobalt(ii) phthalocyanine
chemically modified electrode
url https://kpfu.ru/amperometric-acetylcholinesterase-biosensor-for_340048.html
work_keys_str_mv AT rrdavletshina amperometricacetylcholinesterasebiosensorforphysostigminedetermination
AT anivanov amperometricacetylcholinesterasebiosensorforphysostigminedetermination
AT gaevtugyn amperometricacetylcholinesterasebiosensorforphysostigminedetermination