Effects of ventilator tubing corrugation design on condensate accumulation

Abstract Background The use of adequately humidified and heated gas during invasive mechanical ventilation (IMV) is mandatory to prevent complications associated with under humidification. However, condensate formation within ventilator tubing remains a persistent challenge that can compromise venti...

Full description

Saved in:
Bibliographic Details
Main Author: Ding Yang Hsu
Format: Article
Language:English
Published: Springer 2025-08-01
Series:Discover Applied Sciences
Subjects:
Online Access:https://doi.org/10.1007/s42452-025-07060-8
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1849226013359210496
author Ding Yang Hsu
author_facet Ding Yang Hsu
author_sort Ding Yang Hsu
collection DOAJ
description Abstract Background The use of adequately humidified and heated gas during invasive mechanical ventilation (IMV) is mandatory to prevent complications associated with under humidification. However, condensate formation within ventilator tubing remains a persistent challenge that can compromise ventilation efficiency and patient safety. This study investigates whether the shape of ventilator tubing influences the formation of condensate, focusing on its implications for both non-invasive ventilation (NIV) and IMV. Methods A commercially available adult heated tubing was used as a control group(conventional heated tubing) alongside nine experimental tubing designs with distinct combinations of corrugation angle (non-skewed, left-skewed, right-skewed) and wavelength (narrow, standard, wide). The tubing sets were connected to a heated humidifier, and condensate accumulation was measured at 1, 2, 4, 8, and 24 h under controlled environmental conditions (23 °C–26 °C, relative humidity 40–60%). Data were averaged across three trials, and a two-way analysis of variance (ANOVA) was conducted to assess the effects of design variables. Results By 8 h, inter-tube differences reached 11.8 g, increasing to 42.8 g by 24 h. Two-way ANOVA revealed that wavelength significantly affected condensate buildup (p = 0.016), while skew angle did not (p = 0.126). However, their interaction was significant (p = 0.012). WN showed the highest 24-h condensate accumulation (265.07 ± 6.99 g), whereas SR had the lowest (222.27 ± 22.55 g), outperforming the control (SN: 238.83 ± 4.39 g). Conclusion These findings underscore the pivotal influence of wavelength on condensate formation. Compared with the conventional SN design, SR reduced 24-h condensate by approximately 16.56 g (~ 7%). Tailoring tubing designs based on wavelength and skew angle can help optimize mechanical ventilation performance and potentially enhance patient safety and comfort by reducing condensate-related complications.
format Article
id doaj-art-b9869b79c8be4851b4e6be08a4bc9b25
institution Kabale University
issn 3004-9261
language English
publishDate 2025-08-01
publisher Springer
record_format Article
series Discover Applied Sciences
spelling doaj-art-b9869b79c8be4851b4e6be08a4bc9b252025-08-24T11:45:28ZengSpringerDiscover Applied Sciences3004-92612025-08-01791810.1007/s42452-025-07060-8Effects of ventilator tubing corrugation design on condensate accumulationDing Yang Hsu0Department of Industrial Design, Ming-Chi University of TechnologyAbstract Background The use of adequately humidified and heated gas during invasive mechanical ventilation (IMV) is mandatory to prevent complications associated with under humidification. However, condensate formation within ventilator tubing remains a persistent challenge that can compromise ventilation efficiency and patient safety. This study investigates whether the shape of ventilator tubing influences the formation of condensate, focusing on its implications for both non-invasive ventilation (NIV) and IMV. Methods A commercially available adult heated tubing was used as a control group(conventional heated tubing) alongside nine experimental tubing designs with distinct combinations of corrugation angle (non-skewed, left-skewed, right-skewed) and wavelength (narrow, standard, wide). The tubing sets were connected to a heated humidifier, and condensate accumulation was measured at 1, 2, 4, 8, and 24 h under controlled environmental conditions (23 °C–26 °C, relative humidity 40–60%). Data were averaged across three trials, and a two-way analysis of variance (ANOVA) was conducted to assess the effects of design variables. Results By 8 h, inter-tube differences reached 11.8 g, increasing to 42.8 g by 24 h. Two-way ANOVA revealed that wavelength significantly affected condensate buildup (p = 0.016), while skew angle did not (p = 0.126). However, their interaction was significant (p = 0.012). WN showed the highest 24-h condensate accumulation (265.07 ± 6.99 g), whereas SR had the lowest (222.27 ± 22.55 g), outperforming the control (SN: 238.83 ± 4.39 g). Conclusion These findings underscore the pivotal influence of wavelength on condensate formation. Compared with the conventional SN design, SR reduced 24-h condensate by approximately 16.56 g (~ 7%). Tailoring tubing designs based on wavelength and skew angle can help optimize mechanical ventilation performance and potentially enhance patient safety and comfort by reducing condensate-related complications.https://doi.org/10.1007/s42452-025-07060-8Respiratory therapyHeated humidifierVentilator tubingCondensate
spellingShingle Ding Yang Hsu
Effects of ventilator tubing corrugation design on condensate accumulation
Discover Applied Sciences
Respiratory therapy
Heated humidifier
Ventilator tubing
Condensate
title Effects of ventilator tubing corrugation design on condensate accumulation
title_full Effects of ventilator tubing corrugation design on condensate accumulation
title_fullStr Effects of ventilator tubing corrugation design on condensate accumulation
title_full_unstemmed Effects of ventilator tubing corrugation design on condensate accumulation
title_short Effects of ventilator tubing corrugation design on condensate accumulation
title_sort effects of ventilator tubing corrugation design on condensate accumulation
topic Respiratory therapy
Heated humidifier
Ventilator tubing
Condensate
url https://doi.org/10.1007/s42452-025-07060-8
work_keys_str_mv AT dingyanghsu effectsofventilatortubingcorrugationdesignoncondensateaccumulation