Validity of the Actigraph-GT9X accelerometer for measuring steps and energy expenditures in heart failure patients.
<h4>Introduction</h4>The ActiGraph (AG) accelerometer is widely used to assess physical activity (PA) in heart failure (HF) patients. However, the validity of the AG in this population remains unexplored.<h4>Objective</h4>Therefore, this study examined the criterion validity...
Saved in:
Main Authors: | , , , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Public Library of Science (PLoS)
2024-01-01
|
Series: | PLoS ONE |
Online Access: | https://doi.org/10.1371/journal.pone.0315575 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | <h4>Introduction</h4>The ActiGraph (AG) accelerometer is widely used to assess physical activity (PA) in heart failure (HF) patients. However, the validity of the AG in this population remains unexplored.<h4>Objective</h4>Therefore, this study examined the criterion validity of the AG-GT9X for measuring step counts (SC) and energy expenditure (EE) among HF patients.<h4>Methods</h4>16 patients with HF with preserved ejection fraction (mean age = 60.3±12.1yrs) completed a total of 41 symptom-limited cardiopulmonary exercise tests on a treadmill across multiple time points (median (IQR) = 2.5 (1.5-3.5)). All participants wore the AG (model: GT9X) on both the right ankle and waist locations during the test. Manually counted steps and indirect calorimetry-derived EE served as criterion measures. AG-derived EE was estimated using six different prediction equations previously developed for waist-worn AG. AG-derived measurements were compared with criterion measurements by calculating correlation coefficients, equivalence tests with two one-sided tests, mean absolute percentage error (MAPE), percentage bias, and Bland-Altman plots using mixed models to account for the nested nature of repeated measures within subjects.<h4>Results</h4>Ankle-worn AG-SC was significantly equivalent to the criterion (p < .05) and had lower MAPE (<10%) compared to the waist location, regardless of PA intensity level. Sasaki-EE was significantly equivalent to the criterion (p < .05), with the lowest percentage bias overall (0.7%).<h4>Conclusions</h4>The ankle-worn AG-SC and Sasaki-EE showed better accuracy among HF patients in laboratory settings. Further research is warranted to cross-validate the results in different settings. |
---|---|
ISSN: | 1932-6203 |