Validating the accuracy of Omron J760 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018) + AMD1:2020
Purpose The accuracy of the Omron J760 electronic blood pressure (BP) monitor for upper arm BP measurement in the adult general population was validated following the AAMI/ESH/ISO (ISO 81060-2:2018) + Amd1:2020 protocol. We expect that this device can be used for home blood pressure monitoring, with...
Saved in:
| Main Authors: | Bihua Chen, Guorong Chen, Hong Jiang, Ya Sun, Jie Chen, Dan Wang, Mingzhi Chen, Jia Shi, Yan Peng, Yimin Cheng, Cheng Yang, Yuanyuan Ding, Jin Su, Ming Liu, Fulai Shen, Yicheng Qiu, Yi. Shen, Qiyun Cao, Jiayu Su |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
Taylor & Francis Group
2025-12-01
|
| Series: | Blood Pressure |
| Subjects: | |
| Online Access: | https://www.tandfonline.com/doi/10.1080/08037051.2025.2469264 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Similar Items
-
MODERN VIEWS ON THE VARIABILITY OF BLOOD PRESSURE
by: V. M. Gorbunov
Published: (2015-12-01) -
Comparison between an adapted protocol for home and ambulatory measurement for evaluating night-time blood pressure behaviour
by: Thamyres Costa Borges, et al.
Published: (2025-12-01) -
The “Inverse” Seasonal Blood Pressure Variability Phenotype
by: V. M. Gorbunov, et al.
Published: (2021-07-01) -
Prevalence of newly diagnosed high blood pressure among female adults aged 18 to 50 years in south western Nigeria
by: Ngozi A. Ukangwa, et al.
Published: (2025-01-01) -
Multiple Modern Methods of Blood Pressure Measurement and the 2018 European Society of Hypertesion Guidelines
by: V. M. Gorbunov, et al.
Published: (2020-03-01)